5. Operating the Instrument
124 IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6 2016-04
5.13 Cuvette Measurements
5.13.1 Cuvette Strip
For performing cuvette measurements, a ‘Cuvette’ strip is necessary
Figure 45: Cuvette strip
For a few applications it might be necessary to combine a microplate
measurement with a cuvette measurement. The i-control software therefore
allows the usage of one cuvette strip and one plate strip within one measurement
script. The cuvette measurement has to be positioned before the microplate
measurement. To perform an accurate microplate measurement, the cuvette door
must not be open. The software therefore does not allow the user to use a ‘Move
cuvette OUT’ strip before the microplate measurement (see also chapter 5.12.3).
5.13.2 Cuvette Movements
The cuvette can be moved in and out with the ‘cuvette in’ and ‘cuvette out’
buttons or by selecting Cuvette in/Cuvette out in the Instrument/Movements
dialog box.
Figure 46: Cuvette ‘out’ and ‘in’ button