4. Gas Control Module (Enhanced)
2016-04 IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6 63
4.5.2 Installation Procedure
The following information details the installation procedure:
IMPORTANT
FOLLOW APPROPRIATE GAS HANDLING PRECAUTIONS AND
SAFETY REGULATIONS WHEN SETTING UP THE CO
2
AND/OR N
2
SUPPLY. READ ALL LABEL INFORMATION AND MATERIAL SAFETY
DATA SHEETS (MSDS) FROM MANUFACTURER OR SUPPLIER.
IMPORTANT
ALWAYS USE A REGULATOR APPROVED FOR THE SPECIFIC GAS
WITH HIGH AND LOW-PRESSURE GAUGES.
For details see chapter 4.7 CO2 and N2 Gas Cylinders.
1. Unpack the gas control module and place it on top or near the instrument on
a flat, rigid surface (see Figure 25, below).
Figure 25