6. Instrument Features
142 IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6 2016-04
6.6 Glow Type Luminescence
CAUTION
SWITCH ON THE INSTRUMENT AT LEAST 15 MINUTES BEFORE
STARTING A LUMINESCENCE MEASUREMENT. SOME COMPONENTS
NEED TO WARM UP TO GUARANTEE STABLE CONDITIONS FOR THE
MEASUREMENT.
Luminescence Detection uses the photon counting technique.
Parameters Characteristics
Wavelength Range 380 – 600 nm
Linear Dynamic Range 6 orders of magnitude
Integration Time/well 100 – 20000 ms
Cross Talk % (black plate) < 0.01 %
Measurement range
> 6 orders of magnitude
8 orders of magnitude
(extended dynamic range)
Attenuation of Light
100 (OD2 attenuation filter),
1 (no attenuation)
6.6.1 ATP Glow Luminescence
Plate Type (number of wells) 96
Total Dispensed Volume[µl] 200
Integration Time/well [ms] 1000
ATP Detection Limit 3 fmol/well
CAUTION
SPECIFICATIONS ARE ONLY VALID WHEN THE INSTRUMENT IS
PLACED IN A LOCATION WITH ILLUMINATION < 500 LUX.
STOP
STOP