6. Instrument Features
142 IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6 2016-04
6.6 Glow Type Luminescence
CAUTION
SWITCH ON THE INSTRUMENT AT LEAST 15 MINUTES BEFORE
STARTING A LUMINESCENCE MEASUREMENT. SOME COMPONENTS
NEED TO WARM UP TO GUARANTEE STABLE CONDITIONS FOR THE
MEASUREMENT.
Luminescence Detection uses the photon counting technique.
Parameters Characteristics
Wavelength Range 380 600 nm
Linear Dynamic Range 6 orders of magnitude
Integration Time/well 10020000 ms
Cross Talk % (black plate) < 0.01 %
Measurement range
> 6 orders of magnitude
8 orders of magnitude
(extended dynamic range)
Attenuation of Light
100 (OD2 attenuation filter),
1 (no attenuation)
6.6.1 ATP Glow Luminescence
Plate Type (number of wells) 96
Total Dispensed Volume[µl] 200
Integration Time/well [ms] 1000
ATP Detection Limit 3 fmol/well
CAUTION
SPECIFICATIONS ARE ONLY VALID WHEN THE INSTRUMENT IS
PLACED IN A LOCATION WITH ILLUMINATION < 500 LUX.
STOP
STOP