6. Instrument Features
2016-04 IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6 141
6.5 Absorbance
Parameters Instrument Type Characteristics
Wavelength Range Infinite F200 PRO 230 – 1000 nm
Infinite M200 PRO 230 – 1000 nm
no filter necessary, selection
in 1 nm steps possible
Measurement
Range
Both 0 – 4 OD
The following specifications are valid for the wavelength range from
300 – 700 nm for Infinite 200 PRO series.
Plate type (number of wells) 96
Accuracy 0 – 2 OD
< ± (1 % + 10 mOD)
Accuracy 2 – 3 OD
< ± 2.5 %
Baseline Flatness
± 10 mOD (1 sigma)
Infinite M200 PRO
Wavelength Accuracy
≤ ± 1.5 nm
λ > 315 nm;
≤ ± 0.8 nm
λ ≤ 315 nm
Infinite F200 PRO
Wavelength Accuracy Dependent on filters used
The specifications are valid for measurements performed with 25 flashes (reads)
per well.