6. Instrument Features
2016-04 IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6 147
6.12 Injector Specifications
Parameters Characteristics
Accuracy < 10% for injection volume of 10 µl
< 2% for injection volume of 100 µl
< 0.7% for injection volume of 450 µl
Precision < 10% for injection volume of 10 µl
< 2% for injection volume of 100 µl
< 0.7% for injection volume of 450 µl
6.12.1 Injector Reagent Compatibility
The injector system of the Infinite F200 PRO and Infinite M200 PRO instruments
consists of the following materials:
• Teflon (PTFE): Tubing, valve plug, seal
• KelF: Valve body
• SC05: Injector needles
Please refer to the following list for reagent compatibility. Rating ‘A’ indicates a
good compatibility with the injector system. Chemicals with a rating ‘D’ must not
be used with the Infinite injectors. They will severely damage the injector system.
‘A’ Rated Chemicals ‘D’ Rated Chemicals
Acetic Acid < 60% Butyl Amine
Acetonytrile Carbon Tetrachloride (dry)
Chloroform Diethyl Ether
Dimethyl Formamide Ethanolamine
Ethanol Ethylene Diamine
Hexane Furfural
Methanol (Methyl Alcohol) Hydrofluoric Acid
Sulfuric Acid, diluted (Concentration
1 N)
Monoethanolamine
Tetrahydrofuran Potassium Hydroxide (Caustic Potash)
Water, Deionized Potassium Hypochlorite
Water, Distilled Sodium Hydroxide
Water, Fresh Sodium Hypochlorite
Concentrated Sulfuric Acid
CAUTION
THE INFORMATION IN THIS TABLE HAS BEEN SUPPLIED TO TECAN
AUSTRIA BY OTHER REPUTABLE SOURCES AND IS TO BE USED
ONLY AS A GUIDE IN SELECTING EQUIPMENT FOR APPROPRIATE
CHEMICAL COMPATIBILITY. BEFORE PERMANENT INSTALLATION,
TEST THE EQUIPMENT WITH THE CHEMICALS AND UNDER THE
SPECIFIC CONDITIONS OF YOUR APPLICATION.