6. Instrument Features
2016-04 IFU for Infinite 200 PRO No. 30052730 Rev. No. 1.6 143
6.7 Flash Type Luminescence
CAUTION
SWITCH ON THE INSTRUMENT AT LEAST 15 MINUTES BEFORE
STARTING A LUMINESCENCE MEASUREMENT. SOME COMPONENTS
NEED TO WARM UP TO GUARANTEE STABLE CONDITIONS FOR THE
MEASUREMENT.
Luminescence Detection uses the photon counting technique.
Parameters Characteristics
Wavelength Range 380 – 600 nm
Measurement Range
> 6 orders of magnitude
8 orders of magnitude
(extended dynamic range)
Integration Time/well 100 – 20000 ms
Cross talk % (black plate) < 0.01 %
Attenuation of Light
100 (OD2 attenuation filter),
1 (no attenuation)
ATP Detection Limit < 80 amol/well
CAUTION
SPECIFICATIONS ARE ONLY VALID WHEN THE INSTRUMENT IS
PLACED IN A LOCATION WITH ILLUMINATION < 500 LUX.
STOP
STOP