EasyManuals Logo

Tektronix 222 User Manual

Tektronix 222
167 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #126 background imageLoading...
Page #126 background image
Performance Specifications-222 DSO Operators
8.4
ENVIRONMENTAL SPECIFICATIONS
Environmental Requirements
Instrument will meet the requirements of Tektronix Standard
062-2853-00, Class 3.
The instruments meets all the following
MIL-T-28800D require-
ments for Type
Ill,
Class 3 equipment, except where noted
otherwise.
Temperature
Operating
-10°C to
+55"C
(+
14°F to
+
131
OF).'
Nonoperating
-51 OC to +71 "C (-60°F to
+
160°F).1
Tested to MIL-T-28800D, para 4.5.5.1.3 and 4.5.5.1.4 except
that in para 4.5.5.1.3, steps
4
and
5
are performed before step
2
(-51
OC nonoperating test). Equipment shall remain off upon
return to room-ambient temperature during step 6. Excessive
condensation shall be removed before operating during step 7.1
Altitude
Operating
4,570 meters (1 5,000 feet). Maximum
operating temperature decreases
1
"C
per 1,000 feet above 5,000 feet.1
Nonoperating To 15,240 meters (50,000 feet).'
Performance Requirement
not
checked in manual.
REV
APR
1989

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 222 and is the answer not in the manual?

Tektronix 222 Specifications

General IconGeneral
BrandTektronix
Model222
CategoryTest Equipment
LanguageEnglish

Related product manuals