Performance Verication
Measure the Oscilloscope Gain
1. Connect a BNC-to-Dual Banana adapter
to the voltage output terminals of the
calibrator.
2. Connect a BNC cable betwee n the adapte r
and the oscilloscope. (Leave the probe
connectedtokeepitwarmedup.)
3. Set the oscilloscope:
Vertical se nsitivity to 500 mV/div
Vertical position and offset to 0.0 V
Horizontal to 400 or 500 us/div, trigger
50%
Acquisition mode to average 16
Input to 1 M a nd coupling to DC
Automeasurement to display
AMPLITUDE
TCP0030 C urren t Probe Instruction Manual 55