Adjustments
Adjustments
The procedures that follow describe adjustments to the probe to bring the performance within the wa rranted sp ecications.
DC gain accuracy
Equipment Required
Refer to the Performance Verication procedure for the equipme nt required. You also need an insulated, at-bladed adjustment tool.
Equipment Setup
Refer to the Performance Verication procedure for the equipment setup.
CAUTION. To avoid ESD da mage to the probe, use an antistatic w rist strap and work at a static-approved workstation when
handling the probe
TCP0030 C urren t Probe Instruction Manual 63