Performance Verication
Measure the Probe Gain
10. Connect the probe to the oscilloscope
channel you used in the previous test.
11. Connect the 5-tu rn DC curre nt loop to the
calibrator.
12. Set the ca librator to AC Amps.
13. Degauss the probe, and th en connect it to
the 5-turn current loop.
14. Set the prob e range to 5 A.
15. Set t he oscilloscope:
Vertical se nsitivity to 500 mA/div
Trigger at 50%
Acquisition mode to average 16
CouplingtoDC
Automeasurement to display
AMPLITUDE
TCP0030 C urren t Probe Instruction Manual 57