5 FAQ
51
• The WiFi data logger is displaying error code E69, what can I do?
1. The Account ID contained in the configuration file is missing or is not
valid. Create a new configuration file and save this on the WiFi data logger.
2. An attempt was made to log the testo 160 E WiFi data logger into the
Cloud without any external probes connected.
Connect the required
external probes before logging in.
• The WiFi data logger (160 IAQ) is displaying error code E75, what can I
do?
A firmware update for the WiFi data logger failed.
The internet connection was interrupted during the transmission or the data
was not received intact by the WiFi data logger for other reasons. Check
whether there is a stable connection from the WiFi data logger to the
access point (WLAN router). Check the access point's internet connection.
The data will be transferred during the next communication cycle.
Alternatively: Initiate data transmission manually by pressing the control key
on the WiFi data logger.
• The WiFi data logger (160 IAQ) is displaying the warning message
Err AccountID, what can I do?
The AccountID contained in the configuration file is not valid.
Use the Quick Start Guide to create a new configuration file and save this
on the WiFi data logger.
• The WiFi data logger (160 IAQ) is displaying the warning message
no AccountID, what can I do?
There is no AccountID in the configuration file.
Use the Quick Start Guide to create a new configuration file and save this
on the WiFi data logger.
• The WiFi data logger is displaying the warning message no License,
what can I do?
The WiFi data logger cannot be logged in because the number of WiFi data
loggers permitted to log in has been exceeded or your testo 160 licence has
expired.
Log off another WiFi data logger, extend or renew your testo 160 licence.
• The WiFi data logger (160 IAQ) is displaying the warning message
not Active, what can I do?
The WiFi data logger has been deactivated. It is not storing, and therefore
not sending, any measurement data to the Testo Cloud.
Activate the WiFi data logger (under Configuration --> WiFi data logger)
when the WiFi data logger needs to store and send measurement data