4. *RST – reset – return instrument to defaults
When the *RST command is sent, the instrument performs the following operations:
· Returns the instrument to default conditions.
· Cancels all pending commands.
· Cancels response to any previously received *OPC and *OPC? commands.
5. *TST? – self-test query – run self test and read result
Use this query command to perform the instrument self-test routine. The command places the
coded result in the Output Queue. A returned value of zero (0) indicates that the test passed,
other values indicate that the test failed and an error code is placed into the error queue.
6. *WAI – wait-to-continue – wait until previous commands are completed
The *WAI command is a no operation command for the instrument and thus, does not need to
be used. It is there for conformance to IEEE488.2.