4.3.4 Error Reporting
The device stores errors in a queue containing up to 10 entries. The error queue may be read
out by the ‘SYSTem:ERRor[:NEXT]?’command. The following table lists all error numbers
and the according descriptive messages. Note: negative numbers are defined by SCPI while
positive error numbers are device dependent.
Error
Description
0
No error
1
The error couldn't be specified more precisely
2
Floating point domain error
3
Device temperature too high
4
General GUI error
5
Authentication required for operation
6
Authentication process failed
7
Operation is not allowed in service mode
8
Operation is allowed in service mode only
9
A measurement is currently in process
14
LED head is missing or it is of unknown type
15
Power supply error
20
Operation not allowed while LED output is on
21
Wrong operating mode for this operation
22
INTERLOCK circuit is open
23
LED is overheated
24
Operation not allowed because of a 'OPEN CIRCUIT' condition
26
VTM module error
27
PRM module error
28
PRM module short circuit detected
29
VTM module overheated
30
PRM module overheated
31
Output current limit reached
32
Sensor failed
33
Supply 3.3V digital failed
34
Supply 1.2V digital failed
35
Supply 12V analog failed
36
Supply -12V analog failed
37
Supply 5V analog failed
38
Supply 12V VTM module and internal fan failed

Questions and Answers: