119Service and Maintenance
Go to the Camera Settings Setup panel and select a small number of
averages.
There is probably the trigger functionality activated but no trigger signal
applied. Check the 'Hardware Trigger' control within the Camera Settings
Setup and see the Trigger Input chapter for correctly applying a trigger
signal.
3. The menu items, toolbar buttons and setting values in the main panel are
not reacting on your clicks
There is a Wavefront Sensor popup window open like 'Wavefront Sensor
Message'. The input focus lies only on this open popup window. Close the
open popup window to get the input focus back to the main panel.
4. Erroneous or no measurement data
Check the correct Microlens Array (MLA) selection in the Wavefront Setup
, probably you have physically installed a different MLA than selected
within the WFS software.