EasyManua.ls Logo

Tonghui TH2523/A - Page 63

Tonghui TH2523/A
86 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TH2523 Series Operation Manual Ver1.2
58
TRIGger [:IMMediate]
:SOURce INTernal
EXTernal
BUS
HOLD
:DELay <value>
MIN
MAX
The :IMMediate command triggers a measurement.
Command syntax: TRIGger[:IMMediate]
For exampleWrtCmd(“TRIG”);
The :SOURce command sets the mode of trigger source.
The :SOURce? query returns the current mode of trigger source.
Command syntax:
INTernal
TRIGger:SOURce EXTernal
BUS
HOLD
Where:
INTernal Be automatically triggered by the instrument and be the defaut setting
of the instrument.
EXTernal Be triggered by the HANDLER interface.
BUS Be triggered by the RS232 interface or the GPIB interface.
HOLD Press the TRIGGER button on the panel.
For example: WrtCmd(“TRIG:SOUR BUS”);
Query syntax: TRIGger:SOURce?
Return format:
INT
EXT <NL^END>
BUS
HOLD

Table of Contents

Related product manuals