TH2523 Series Operation Manual Ver1.2
68
The : STATIstics:NORmalA command sets the nominal value of primary parameter in
statistical function.
Command syntax:
STATIstics: NORmalA <value>
For example: STATIstics: NORmalA 50
Set the nominal value of primary parameter to 50, the unit can be Ω or V according to the
selected measurement parameter.
Query syntax: STATIstics: NORmalA?
Response: < NR3><NL^END>
The : STATIstics:NORmalB command sets the nominal value of secondary parameter in
statistical function.
Command syntax:
STATIstics: NORmalB <value>
For example: STATIstics: NORmalB 50
Set the nominal value of secondary parameter to 50, the unit can be Ω or V according to the
selected measurement parameter.
Query syntax: STATIstics: NORmalA?
Response: < NR3><NL^END>
The : STATIstics:DEViation? query returns the statistical deviation.
Query syntax: : STATIstics:DEViation?
Response: <σn(NR3)>
For example: Syntax: :STATI:DEV?
Response: 0.0159E-3
The : STATIstics:SampleVARiance? query returns the statistical variance.
Query syntax: STATIstics: SampleVARiance?
Response: <sn(NR3)>
For example: Syntax: :STATI:DEV?
Response: 0.0159E-3
The : STATIstics:CP? query returns the Process Capability Index (CP).
Query syntax: : STATIstics:CP?