TH2523 Series Operation Manual Ver1.2
70
The :TRACe:AM<max value>, <min value> command is used to set the A Max value and the
A Min value in trace scanning setting.
For example: :TRACe:AM 3.1600E+03,2.0000E+02
Query syntax: :TRACe:AM?
Response: <max value NR3>, <min value NR3>
The :TRACe:BM<max value>, <min value> is used to set and return the BMax value and the
BMin value in the trace scanning setting.
For example: :TRACe:BM 5.1600E+01,1.0000E+01
Query syntax: :TRACe:BM?
Response: <max value NR3>, <min value NR3>
The :TRACe:ASTOP1 OFF
<value>
:TRACe:ASTOP2 OFF
<value>
:TRACe:BSTOP1 OFF
<value>
:TRACe:BSTOP2 OFF
<value> commands are used to set or return the stop point of trace
scanning.
Command syntax:
:TRACe:ASTOP1 OFF Turn off the judging function of this stop point.
:TRACe:ASTOP1 <value> Set the value of the stop point as value.
Query syntax::TRACe:ASTOP1?
Response: OFF or <value NR3>
The :TRACe:SCAN STARt (valid only in TRACE DISP page)
STOP command is used to start or stop trace scanning or to return the
current state of trace scanning.
Command syntax:
:TRACe:SCAN STARt
:TRACe:SCAN STOP
Query syntax: