18
PREPARATION BEFORE USE
SETTING THE INSTALLING DISTANCE
You can change the installing distance to 60cm, 80cm, 95cm, 1m and 1.5m*.
* Only when selecting the H1 type.
According to the installing distance, it is necessary to change the DIP switch of the instrument.
1 To perform setting for the instrument, open the cover at the rear of the sensor detector unit.
*The installing distance is set to 80cm as the default when shipped.
CAUTIONS
To avoid electric shock, when setting the installing distance do not
make any contact with the patient and the sensor detector unit at the
same time.
Before setting the installing distance, make sure that the power switch
of the instrument is OFF. If it is ON, you may receive an electric
shock.
NOTES
To change the installing distance setting, ask your service personnel.
NOTES
To change the setting of the instrument, ask your service personnel.
Installing distance
Instrument
SW01-1 SW01-2 SW01-3
60cm OFF OFF OFF
80cm OFF ON OFF
95cm, 1m, 1.5m OFF OFF ON