5. Preparation and Functions
22
G
Recharging should take place in a room with an ambient temperature range of 10°C to 40°C
(50°F to 104°F).
Do not perform charging with others except the AC/DC converter AD-15.
For longer battery life, conform to the suggested charging time to the extent possible.
The battery source will discharge when stored and should be checked before using with
instrument.
Be sure to charge stored battery source every 3 or 6 months and store in a place at 30 °C or
below. If you allow the battery to become completely discharged, it will have an effect on future
charging.