8. Check and Adjusting
48
The screen will be in the X-axis check and adjusting mode.
Auto leveling on the RL-SV2S is complete and the laser will emit.
4 Turn the LS-80L to the high precision mode.
5 Detect the center of the laser beam on the wall with the LS-80L and mark it. (X1)
6 Press the key.
7 Loosen the centering screw and rotate the RL-SV2S 180°, and tighten the screw to secure.
The RL-SV2S X+ surface will face the wall. When rotating the RL-SV2S, ensure that the
instrument height is not misaligned.
The RL-SV2S auto leveling is complete and the laser will emit.