7 Press the [ENT] key on the RC-300W. The unit will
rotate 180 degrees internally and start auto-
leveling. The RC-300W will now display “X:2”.
8 When the rotary head starts rotating, do not move
the LS-70A laser sensor to locate the beam center.
Instead, note the beam location as shown on the
laser sensor display. If one of the three center LCD
indicators is lit (center bar, Down arrow-point, Up
arrow-point) calibration is OK and no adjustment is
needed. Repeat steps 4-8 for the remaining axis (Y
in this example).
If adjustment is necessary proceed to the
“Adjusting” instructions below before checking the
remaining axis.
X:2
Press [ENT]
The unit turns to 180° and starts auto-
leveling.
Note the position of the X2 laser
beam on the LS-70A display. If
one of the three center indicators
is lit, no calibration adjustment is
necessary.