EasyManua.ls Logo

Topcon TRC-NW400 - Electromagnetic Compatibility

Topcon TRC-NW400
106 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
80
SPECIFICATIONS AND PERFORMANCE
ELECTROMAGNETIC COMPATIBILITY
This product conforms to the EMC standard (IEC 60601-1-2 Ed.3: 2007).
a) MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the ACCOMPANYING
DOCUMENTS.
b) Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
c) The use of ACCESSORIES, transducers and cables other than those specified, with the exception of
transducers and cables sold by the manufacturer of the EQUIPMENT or SYSTEM as replacement
parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
EQUIPMENT or SYSTEM.
d) The EQUIPMENT or SYSTEM should not be used adjacent to or stacked with other equipment. If adja-
cent or stacked use is necessary, the EQUIPMENT or SYSTEM should be observed to verify normal
operation in the configuration in which it will be used.
e) The use of the ACCESSORY, transducer or cable with EQUIPMENT and SYSTEMS other than those
specified may result in increased EMISSION or decreased IMMUNITY of the EQUIPMENT or SYS-
TEM.
Item Article code Cable Shielded Ferrite Core Length (m)
AC power cable (for TRC-NW400) 44804 70150 No No 1.5
LAN cable 47005 35100 No Yes 2.0
USB mouse cable (for TRC-NW400) - Yes No 1.8
USB memory - - - -
Personal computer 47010 77200 - - -
USB keyboard cable - Yes Yes 2.0
USB mouse cable - Yes No 1.8
AC power cable (for personal computer) - No No 2.0
Personal computer monitor 47010 77300 - - -
AC power cable (for monitor) - No No 1.8
DVI cable - Yes Yes 2.0
Insulation transformer box 47010 77000 - - -
AC power cable (for transformer) 44804 70150 No No 1.5
FG cable - No No 2.0
Guidance and manufacturer's declaration - electromagnetic emissions
The TRC-NW400 is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the TRC-NW400 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The TRC-NW400 uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equip-
ment.
RF emissions
CISPR 11
Class B
The TRC-NW400 is suitable for use in all establishments
including domestic and those directly connected to the pub-
lic low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Complies

Table of Contents