SPECIFICATIONS AND PERFORMANCE
Guidance and manufacturer's declaration - electromagnetic immunity
The TRC-NW400 is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the TRC-NW400 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150kHz to 80MHz
3 V/m
80MHz to 2.5GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the TRC-NW400, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 80MHz to 800MHz
d = 2.3 800MHz to 2.5GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the rec-
ommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site sur-
vey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1
NOTE 2
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
b
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be pre-
dicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF trans-
mitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the TRC-NW400 is used exceeds the applicable RF compliance level above, the
TRC-NW400 should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the TRC-NW400.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
P
P
P