Feedback Controlyzer DigiLock 110
Page 52
Status: 5.12.17
Table 4 Input signal path settings
Name Description Standard Value
Input offset Offset to be subtracted from the input signal
before digitalization to reduce the signal offset
for further analog amplification
0.0 V
Gain Analog gain to amplify the signal to take advan-
tage of the maximum input range of the analog-
to-digital converter (ADC) and, hence, its resolu-
tion +/-2 V
1
Invert Inverts the input signal
1
1. The signal can also be inverted by using the controller polarity, i.e the inversion is mostly needed when using the analog
path also.
off (box not checked)
High-pass filter High-pass filter of given frequency and order bypass (box checked)
Low-pass filter Low-pass filter of given frequency and order bypass (box checked)