5.4 mV Calibration Messages
1. If a mV calibration has been successfully performed, the smartCHEM-Titro will display the
following message and the offset of the sensor.
Calibration OK
Offset=-10.0mV
2. If a mV calibration has failed, the smartCHEM-Titro will display the following message and
the failed offset value of the sensor.
Calibration Failed,
Offset=-100.0mV
3. The smartCHEM-Titro has an allowable Offset range of –60 to +60 mV. If calibration fails
due to the Offset being outside these limits, then please consult the Troubleshooting guide
(section 17.2) for possible remedies.