15. Initialising the smartCHEM-Titro
If the calibration settings of the smartCHEM-Titro exceed the allowable limits, the unit may need
to be initialised to factory default values. This action may be required if a sensor is replaced.
Initialising the unit will reset all user setups to their factory default values and erase all readings
logged in memory. Ensure that any valuable data has been noted or downloaded and saved
before proceeding.
To initialise the smartCHEM-Titro…
1. Select Initialise from the Setup menu (
→
→→
→ F4:Setup →
→→
→ F5:Initialise
)
2. The smartCHEM-Titro now asks if you are sure that you wish to initialise ALL parameters, or
just the Slope B from a 3 point pH calibration. The latter is only offered when a 3 point pH
calibration has actually been performed…
Initialise
F1:All
F3:pH SlopeB
Press to initialise ALL parameters, reset all factory default settings and clear the memory.
Press to clear just the Slope B from a 3 point pH calibration.
3. When returning the meter to normal display mode, note that the decimal points have been
replaced with a “
∗
∗∗
∗
”, to indicate that each parameter requires re-calibration.
16. Instrument firmware version number
If you need to contact TPS for any further technical assistance, the version number of your
smartCHEM-Titro firmware may of benefit to us. The version number is displayed by the
smartCHEM-Titro at turn-on.