4.6 pH Calibration Messages
1. If a 1-point calibration has been successfully performed, the smartCHEM-Titro will display
the following message and the asymmetry of the electrode.
Asymmetry
Calibration OK
+0.10pH Asym
2. If a 1-point calibration has failed, the smartCHEM-Titro will display the following message
and the failed asymmetry value of the electrode.
Calibration Failed,
Repeat Cal. or
Initialise
1.1 pH Asymmetry
3. If a 2-point calibration has been successfully performed, the smartCHEM-Titro will display
the following message and the asymmetry and slope of the electrode.
Slope & Asymmetry
Calibration OK
+0.10pH Asym
99.0% Slope
4. If a 2-point calibration has failed, the smartCHEM-Titro will display the following message
and the failed slope value of the electrode.
Calibration Failed,
Repeat Cal. or
Initialise
80.0% Slope
5. If a 3-point calibration has been successfully performed, the smartCHEM-Titro will display
the following message and the asymmetry and second slope of the electrode.
3 Point
Calibration OK
99.0% SlopeB