23
If the calibration was not successful, one of the following error messages will appear.
"Err0" - The calibration test load or the keyed-in load is larger than the full capacity of the
instrument. Change the calibration test load or check the input data.
"Err1" - The calibration test load or the keyed-in load is smaller than 1% of the full capacity
of the instrument. Change the calibration test load or check the input data.
"Err2" There is not enough signal from the force sensor to complete the calibration
process. Most common causes include incorrect force sensor wiring, a mechanical ob-
struction or a faulty (damaged) force sensor.
Take the indicated action to correct the problem, and then perform a new calibration.