When all connections have been made, tighten the cord grips, re-install the rear cover, and then
tighten the acorn nuts.
Load Cell Connections
Connect your shielded load cell cable to the load cell terminal block using the table below.
Load Cell Terminal Block (J1)
Marking
Function
Marking
Function
SH
Shield
E-
- Excitation
S-
- Signal
SEN+
+ Sense
S+
+ Signal
E+
+ Excitation
SEN-
- Sense
Load Cell Terminal Block
From the Factory, the indicator is configured for 4-wire load cells. If you wish to utilize Sense
wires, then you must remove the jumper blocks from R47 and R48, which are located adjacent
to the Load Cell Terminal Block (J1), circled in red below: