Input the number of asphalt samples (the same number of
samples used with the “master” gauge).
You have the option of manually entering the count data using
the keypad or automatically taking the measurements using the
gauge.
To manually enter sample data, press 〈1〉. The samples
must have been previously measured in either the
Calibration or Cross Calibration mode.
For gauge derived cross calibration data collection, press
〈2〉.
This procedure assumes the gauge derived method.
1. The gauge will request a background count. If the
background count is not current or the gauge has been
moved, press
〈YES〉. Empty the sample chamber and press
〈START〉.
2. If the background count is acceptable, press 〈YES〉.
3. Input the sample %AC. The display requests the first
sample.
4. To measure the first sample, press 〈START〉. After the
measurement is complete, the gauge requests that the
remaining samples be measured.
Fi el d Gauge
Measur ement
1-