To print the data, press 〈3〉. Continue by pressing 〈1〉. The
gauge will request a transfer number.
Input the transfer number (up to 8 numbers). This number can
be composed of the gauge serial numbers. For example, if the
“master” gauge serial number is 650 and the “field” gauge serial
number is 658 the transfer number can be 650.658. After
entering the transfer number, press
〈ENTER〉.
The cross calibration is complete and will be stored.
At this time the “master” gauge can be used to perform
calibrations for different mix designs. Calibrations may now be
transferred to the “field” gauges without additional “field” gauge
calibrations.
Transfer
Once a cross calibration has been performed and stored,
“master” gauge calibrations may be transferred to the “field”
gauge without the need for a complete “field” gauge calibration.
On the “Master” Gauge
Prepare asphalt samples for the particular mix being used and
perform a calibration (see Chapter 4) on the “master” gauge.
After completion of the calibration, record the sample weight, the
“master” gauge background count, and calibration constants A1,
A2 and A3.