Model 3440
4–3
After the gauge completes its count time, the display will be:
The reading may be stored for later use by pressing the STORE key.
Refer to page
4–4.
ASPHALT MODE
Enable the Asphalt mode prior to taking a measurement. Refer to Chapter 2
for more details on gauge setup.
Ensure that the asphalt surface has been properly prepared. Refer to
Chapter 3 for more detail in site preparation. Place the gauge at the proper
location and move the source rod to the backscatter position (refer to
Figure 21 on page 2–14).
NOTE
In some applications, the source rod may be extended to place
the gauge in a Direct Transmission mode for use on soft
asphalt.
Verify that all parameters have been set to the correct values.
Start a measurement by pressing START/ENTER. The display will be:
After the gauge completes its count time, the display will be:
The above display will remain until a function key is pressed, a new
measurement is taken, the
NO/CE key is pressed, or the gauge shuts off
due to inactivity.
The reading may be stored for later use by pressing the STORE key.
Refer to page
4–4.
%MA= xxxx
%
WD= xx
M= xxxx %M= xx
%VOIDS= xxxx
%PR=
xxx %
DD= xxx
WD= xxx
M= xx %M= xx
Depth: xx
MA
:
xxxx
Ti
me
:
xxx
se
c.