Moisture Offset – Gauge Derived
From the last display on the previous page, select 〈1〉 for the display:
For a true gauge-derived offset, two values must be input: the true
moisture value, which is determined by laboratory analysis, and the gauge
moisture, which is the gauge reading on the material to be tested.
If the above values are correct, press 〈START/ENTER〉.
To input a new true moisture value, press 〈1〉 for the display:
Input the laboratory value and press 〈START/ENTER〉.
To input a new gauge moisture value, press 〈2〉 for the display:
KEYPAD ENTRY. To use the keypad to enter the gauge moisture reading
from a previous test, press
〈1〉. The display is:
1-
True
M
x.x
%
2-
Gauge
M
x.x
%
ENTER
to
enable
x.x
%
Press ENTER
wh
en
c
o
mp
l
et
ed
Select:
1-
Keypad entry
2-
Me
asu
rem
en
t
x.x
%
Press ENTER
when completed