The compactor raises the loading head to the “home” position. The
screen is:
If you do not want to verify the calibration, press 〈NO〉. To print the
calibration results, press 〈YES〉 at the print calibration prompt.
Open the chamber door and remove the standard. The control unit
returns to the Calibration menu. To return to the Machine Idle
screen, press 〈ESC〉.
To verify the height calibration, press 〈YES〉. The control unit again
requests the standard height. Enter the standard height to the nearest
hundredths and press 〈ENTER〉. Press 〈RAM DOWN〉.
The compactor lowers the loading head to the standard. Check the
compactor measured height displayed on the control unit, then press
〈ENTER〉.
The compactor raises the loading head to the “home” position and
allows the user to verify the sample height again. To verify the
sample height again, press 〈YES〉 at the prompt and repeat the
above procedure.
To exit the height calibration, press 〈NO〉 at the verify prompt. The
control unit returns to the Calibration screen. To return to the
Machine Idle screen, press 〈ESC〉.
Open the chamber door. Remove the standard, specimen papers, and
puck.
Do You Want To
Verify
Another
H
eigh
t