Model 4140
3-7
The compactor raises the loading head to the “home” position. The
screen is:
If you do not want to verify the calibration, press NO. To print the
calibration results, press YES at the print calibration prompt.
Open the chamber door and remove the standard. The control unit
returns to the Calibration menu. To return to the Machine Idle
screen, press ESC.
To verify the height calibration, press YES. The control unit again
requests the standard height. Enter the standard height to the nearest
hundredths and press ENTER. Press RAM DOWN.
The compactor lowers the loading head to the standard. Check the
compactor measured height displayed on the control unit, then press
ENTER.
The compactor raises the loading head to the “home” position and
allows the user to verify the sample height again. To verify the
sample height again, press YES at the prompt and repeat the
above procedure.
To exit the height calibration, press NO at the verify prompt. The
control unit returns to the Calibration screen. To return to the
Machine Idle screen, press ESC.
Open the chamber door. Remove the standard, specimen papers, and
puck.
HEIGHT CALIBRATION
Do You Want To
Verify
Another
H
eigh
t
3. CALIBRATION