Using the UP/DN arrow keys raise the value of
“R” if the display reads high. Lower the value
of “R” if the display reads low.
(Example: if the display reads 2 psi, then raise
the value of “R” by 2; if it reads 2Ih then lower
the value of “R” by 2)
Press the program key until PROGRAM WAIT
is displayed. The autoclave will recalibrate and
return to the Ready state.
If the display now reads 00, with the door of the
autoclave open, then continue on to sec 5.4.3.2”
5.4.3.2 Collecting pressure data
Select a program with a sterilization
temperature of 273 °F (134 °C), make sure the
door is closed securely and press START.
Let the unit cycle to maximum pressure that will
be reached once the unit has entered the
sterilization mode.
Record the temperature and pressure from the
digital display.
Abort the cycle by pressing the STOP key.
Consult the Saturated Steam Table for the
Absolute pressure that corresponds to the
temperature recorded in the previous step and
record that value.
Add 14.5 to the pressure reading from the
digital display and record this value
5.4.3.3 Entering pressure data
Access the calibration mode as described in para
5.4.4.
The cursor will appear flashing under the value
for “A” in the upper row.
— Press the PROGRAM key to advance to the next
value.
— The
cursor on the upper row will move one
position to “R”.
— Press the PROGRAM key to advance to the next
value that will be in the bottom row.
— The cursor will be positioned under the value for
“A” in the bottom row.
Enter the value for “A”. This will be the value
of absolute pressure recorded from the Saturated
Steam Table