8 SOFTWARE PROGRAMMING PARAMETERS
8.1 General
The software version that runs the control system of the EHS autoclave
can be found by turning the unit on and watching the display. The
version number is also printed if a printer is installed.
The software contains a table of parameters of which some of them
define the autoclave, and some of them define the processes in the
autoclave. This section of the manual describes the parameters and how
they control the software.
8.2 Changing Parameters
An access code is needed to change any parameters other than the date
and time.
Two access codes are currently available and each one gives a different
level of access.
Access code 1 – This is the first level of access. This code enables
changing of the sterilization time, sterilization
temperature and the drying time. This level of access
is suitable for a senior office person in the medical
facility.
Access code 13 – This is the second level of access. This code is for the
technician and allows access to make changes such as
calibrations.
To change the parameters listed below, proceed as follows:
a. Select the program cycle, to be modified, by pressing the
SEL.CYCLE key.
b. Press the PROGRAM key until CODE: ∅∅∅ appears
c. Using the UP/DN keys enter the access code for the level of access
desired. (i.e. CODE:013)
d. Select the parameter to be modified, by pressing the PROGRAM
key until the parameter name appears on the display.
e. Set the desired data by means of the UP/DN keys.
f. Pressing the PROGRAM key will enter the modified data into
memory, and move to the next parameter.
g. When finished press the PROGRAM key until the message
PROGRAM WAIT appears. The software is reprogramming itself
with the modifications that were made.
Note: Each of the 6 program cycles has a set of the following
parameters. Changing a specific parameter in one program will
not change it any other program.
Listed below are all the available parameters for the EHS. Each section
describes the parameter, shows the access code required to be able to
make modifications, it shows the minimum and maximum allowed
values and the increments (resolution) by which these values can be
changed. Also included are the pre-set values of the parameters for
each cycle.
Note: If a parameter is modified the only way to return to the original value is
to manually reenter it.