37
8.2.1 SteTemp – Temperature required for sterilization
This parameter will set the desired temperature for sterilization
Access Code - 1
Resolution - 1ºF
Minimum value - 250ºF
Maximum value - 280ºF
Cycle
1 2 3 4 5 6
Default
Values
Value
273 273 250 273 273 -
8.2.2 Ster Time – Time required for sterilization
This parameter will set the time desired for sterilization
Access code - 1
Resolution - 1 minute
Minimum value - 3 minute
Maximum value - 59 minutes
Cycle
1 2 3 4 5 6
2540
3.5 8 30 4 3.5 -
Default
Values
Value
3870
3.5 8 30 8 3.5 -
8.2.3 Dry Time – Time required for drying
This parameter will set the time desired for drying
Access Code - 1
Resolution - 1 minute
Minimum Value - Ø minutes
Maximum Value - 59 minutes
Cycle
1 2 3 4 5 6
Default
Values
Value
0 20 20 20 2 -
8.2.4 HeatT.O. – Sterilization temperature time out
This parameter will set the maximum time allowed, from the
beginning of the cycle, for the chamber to reach sterilization
temperature. When the HeatT.O is exceeded, the program aborts
with the message ‘Low Heat’.
Access Code - 13
Resolution - 30 seconds
Minimum Value - 1200 seconds
Maximum Value - 3180 seconds
Cycle
1 2 3 4 5 6
Default
Values
Value
3000 3000 3000 3000 3000 0