40
8.2.11 VacTime2
Waiting time at the end of each of the remaining
pulses
This parameter sets the wait time at the end of each of the
remaining vacuum pulses, during the prevacuum stage. After
reaching VacDip2, there is a waiting time to allow the unit to
stabilize. This time is defined as VacTime2 and during this time
the vacuum system continues working.
Access Code - 13
Resolution - 1 second
Minimum Value - 3 seconds
Maximum Value - 1800 seconds (30 minutes)
Cycle
1 2 3 4 5 6
Default
Values
Value
3 3 3 3 3 3
8.2.12 WtrResTime – Timed extension for pumping mineral free
water
This parameter sets the length of time to continue pumping
mineral free water into the steam generator after the upper
electrode senses water.
This is necessary to prevent a situation where the water pump
would cycle on and off as the water level rises and falls around
the upper water-sensing electrode.
Access Code - 13
Resolution - 1 minute
Minimum Value - Ø minutes
Maximum Value - 90 minutes
Cycle
1 2 3 4 5 6
Default
Values
Value
5 5 5 5 5 5