42
8.2.15 ReqPrs+ - Overrides the pressure calculated for sterilization
This parameter is used to insure that during sterilization the
temperature is correctly maintained. It does this by raising the
value of the sterilization pressure necessary to maintain a
sterilization temperature. The sterilization pressure is calculated
by the unit based upon the SteTemp parameter.
For example; if the required temperature is 250ºF, the required
pressure calculated by the program is 16 psig.
This pressure can be increased by changing the value of
ReqPrs+. If this value is set to ø, the system will be maintained
at 16 psig during the sterilization stage if this value is set to 1.5,
the system will be maintained at 17.5 psig in the sterilization
stage.
Access Code - 13
Resolution - 0.1 psi
Minimum Value - Ø
Maximum Value - 3 psi
Cycle
1 2 3 4 5 6
Default
Values
Value
0.8 0.8 0.6 0.8 0.8 0