17
8.2.3 Heat Standby – Heating elements operation in Standby Mode
Access Code – 1
Resolution – 1
Value – 0 or 1
— If Heat Standby = 0, the heating elements will not
operate in the Standby Mode.
— If Heat Standby = 1, the heating elements will operate in
the Standby mode, in a shoot mode of 1 second on, 59
seconds off.
8.2.4 Dry Time – Required drying time for the process
Entry Code – 1
Resolution – 1 minute
Minimum Value – Ø minutes
Maximum Value – 99 minutes
Cycle
1 2 3 4 5
E, EK
0 0 30 30 0
Set value
Value
per
model
EA, EKA
1 1 30 30 0
8.2.5 Heat T.O. - Maximum time allowed for the heating stage
(testing)
After this time, if the system does not enter the sterilization stage,
it will receive a ‘Low Heat’ message and the process will abort.
Entry Code – 11
Resolution – 1min
Minimum Value – 20min
Maximum Value – 120 min
Cycle
1 2 3 4 5
Set value
Value
60 60 60 60 90
8.2.6 Ex Mode – The method for exhausting the steam at the end
of the process
Entry Code – 11
Resolution – 1
Ex Mode
Fast Ex 1
N.A. 2
Slow Ex (Waste) 3
Slow Ex (Liquids) 4
Cycle
1 2 3 4 5
Set value
Value
1 1 1 1 4