8.2.7 End Temp - The ending temperature of the process
If at the end of the process the temperature is higher than this
temperature, the process will not end and the door will remain
closed.
Entry Code – 11
Resolution – 1ºC
Minimum Value – 40ºC
Maximum value – 137ºC
Cycle
1 2 3 4 5
Set value
Value
125 121 125 134 95
8.2.8 Air Temp – Controls the closing of the air valve
Defines the temperature of the chamber’s outlet at which the
valve closes.
Access Code - 11
Resolution - 1
Minimum Value - 50°C
Maximum Value - 99°C
Factory pre-set 90°C
Cycle
1 2 3 4 5
Set value
Value
90 90 90 90 90
8.2.9 Calib Code – Calibration Code
This allows entry to the calibration mode.
It must be set on 107 in order to allow the code to be entered.
Entry Code – 11
Resolution – 1
Minimum Value – Ø
Maximum Value – 255
Cycle
1 2 3 4 5
Set value
Value
100 100 100 100 100