96
Error Opening Type file: File Name
Description: This message may be displayed during download process, when
program trying to open ADMC compatible type file (*.mdb) to get from him
parameters data.
Reason: The file may be damaged or used by another application.
Tips: Try to download other file. If operation success, try to find what
application using the file and in the worse case, restart the PC. If the problem
continues, probably the files are damaged. In this case try to replace the file
with new same file from the ADMC installation CD under 'Support' directory.
Invalid Actual High value!
Invalid Actual Low value!
Invalid Read High value!
Invalid Read Low value!
Description: This message displayed when user trying to calculate Gain and
Offset by pressing on 'Calculate' button on 'Calibration' window.
Reason: In one of the text boxes inserted wrong number, or not numeric data.
Tips: Type Numeric data type in the text box.
The Actual High value is not in range!
The Actual Low value is not in range!
The Read High value is not in range!
The Read Low value is not in range!
Description: This message displayed when user trying to calculate Gain and
Offset by pressing on 'Calculate' button on 'Calibration' window.
Reason: In one of the text boxes inserted wrong number. The number must be:
1>Number>1000.
Tips: Type In Range number.
Read High must be equal or great than Read Low!
Description: This message displayed when user trying to calculate Gain and
Offset by pressing on 'Calculate' button on 'Calibration' window.
Reason: Read High value is lees than Read Low value
Tips: Read High value must bee great than Read Low value.