www.shanghaidelan.com
97
Can not calculate. Invalid values found!
Description: This message displayed when user trying to calculate Gain and
Offset by pressing on 'Calculate' button on 'Calibration' window.
Reason: Read High value - Read Low value = 0
Tips: Read High value must bee great than Read Low value.
The Gain value must be equal or les than
The Gain value must be equal or great than
The Offset value must be equal or les than
The Offset value must be equal or great than
Description: This message displayed when user trying to download Gain and
Offset by pressing 'Download' button on 'Calibration' window.
Reason: Gain or Offset value displayed in Gain or Offset text boxes are not in
limits.
Tips: The value in the Gain or Offset boxes must be in Min/Max limits.
Invalid Gain Value!
Invalid Offset Value!
Description: This message displayed when user trying to download Gain and
Offset by pressing 'Download' button on 'Calibration' window.
Reason: Probably the Gain or Offset value displayed in Gain or Offset text
boxes are not numeric data.
Tips: Type numeric value in Gain or Offset text boxes.
Illegal Port number:
Description: This message displayed when user pressing on 'OK' button on
'Options' window.
Reason: Wrong port number is inserted.
Tips: Type value: 0>VAL>17 in 'Port Number' text box.