a few minute drain in order to eliminate any impurities, air
bubbles and/or anything else could damage the machine
andobstructitslers.
- Connectthecoldandpuried(ifavailable)waterhoses,exit-
ing the machine with their respective network connections, as
shown on the installation diagram. It will be the responsibility
of the installer to make sure that the temperature of the cold
water supply is not higher than 5/15°C, otherwise proper
washing of materials cannot be ensured. The connections for
coldandpuriedwatermustnotfeedanyequipmentother
than the instruments washer. During the washing cycle, this
is necessary to prevent the subdivision of the water supply
with other users, thus leading to a substantial increase in
thetimerequiredtollthetank(inthiscaseanalarmwillbe
triggered to alert the user that the maximum time allowed for
water loading is exceeded).
- Ifthemachineisprovidedwithapuriedwaterfeedingsys-
tembuttheplantisnotequippedwithit,thecoldandpuried
water hoses should be connected together.
- The machine can be equipped with a built-in softener, which
has the function of reducing the scale in the water supplied
- Connect the exible hoses to the machine valves posi-
tioned in view at the back, making sure to connect them
correctlybasedonthesalesconguration.
Makesuretoconnectthewaterexiblehosesintheposi-
tionsshowningure8.
• Connectthecoldwaterexiblehosetoconnection1ing-
ure 8.
• Connectthepuriedwaterexiblehose(ifany),toconnec-
tion4ingure8.
WARNING. The ow pipes must not be shortened or
damaged.
The water supply taps must be capable of quickly stopping the
water line, therefore, they must be