5.4.3 Regeneration rooms
The Thermodisinfector is supplied without regeneration salt
which will feed the device automatically during the regener-
ation process.
The salt must be introduced through the tank outlet inside the
tank (4 Fig. 9). To introduce the salt, unscrew the cap (coun-
ter-clockwise)andllthetankwithsalt,payingattentionnotto
spread salt outside the bowl then close the tank with the cap.
Youmustllthesalttankeverytimethemessage“Rellsalt”
is displayed.
- Use only appropriate coarse-grain salt.
- Do not use kitchen salt, crushed salt tabs or other types
different from those indicated, as it may contain insolu-
ble substances.
- Do not pour cleaning liquids or other solutions in the
tank salt.
Failure to comply with these recommendations can lead to a
malfunction of the water softening device. Just before the salt
runs out completely, the display will show the message «Lack
of salt». At this point, the salt should be topped-up as soon as
possible, otherwise an error message will appear and it will no
longer be possible to activate a new cycle if not by resetting.
Proceedasfollowstorellthesalt:
1) Delete the message on the display by pressing the RESET
key for 5 seconds.
2) Open the door and remove any already inserted load.
3) Loosen the salt tank cap and insert the funnel.
4) Intherstlling,introduce½litreofwatertodissolvethe
residue salt. FIRST FILLING ONLY
5) Fill with salt up to the edge. The tank contains about 1 kg
of salt.
6) Thoroughly clean the edge of the tank.
7) Tighten the cap again.
8) Start the «Rinse» cycle.
⇒