Maintenance
WARNING. Do not spray the machine or near it with water
jets, or with pressure devices.
Proceedasfollowstocleantheltersinthewashtank:
- Open the door and remove the load basket.
WARNING: Very hot surfaces.
DANGER of injury: Be careful with sharp and pointed
objects, which are withheld in the baskets.
-Removethecentrallter(1Fig.18).
-Removethemeshlter(2Fig.18).
-Removethetankbottomlter(3Fig.18).
-Carefullycleantheltersandremovetheresidualsubstances.
- Remove the deposits from the drain and clean the drain.
Attheendofthecleaningoperations,rettheltersinsuc-
cession.
Clean the spray arms as follows:
- Open the door and remove the load supports.
WARNING: Very hot surfaces.
- Using the tools provided, loosen and remove both spray arms.
- Rinse the spray arms thoroughly.
- Reassemble and tighten in position.
- Reassemble the spray arms.
7.1.2 Extraordinary maintenance
Extraordinary maintenance operations are not foreseen by
the user but must be performed exclusively by the Tuttnauer
Europe b.v. technical assistance or by an authorised and qual-
iedtechnician.
Extraordinary maintenance operations must be carried out
every 1000 hours of operation or every 12 months