45
Using the machine
6.2 Before use
The Tuttnauer Europe b.v. thermodisinfector can be used for
cleaning and disinfecting:
- Medical instruments.
- Keys, trays, containers.
- Hollow instruments, e.g. suction cannulas, xing them to
the appropriate supports, using the suitable adapters.
6.2.1 Emergency release
In the event of a blackout, or any other need in which it be-
comesdifculttoopenthethermodisinfectordoor,thereisa
manual emergency release that can only be activated if the
door cannot be opened normally.
HAZARD. If the emergency release is used while a pro-
gram is running, very hot water and chemicals may leak.
Therefore, there is a danger of burns, scalding and irri-
tation.
To access the emergency release:
- open the lower door of the machine with the key;
- remove the rubber cap (4 Fig. 13);
- a plastic cable can be seen inside the hole.
Simply pull this cable manually and the door is released.
Re-insert the cap after the operation.
In case of program interruption, proceed with a new treatment
of the load.