5) Start the machine using the ON-OFF switch (1 Fig. 13).
6) Atthebeginningoftherstcyclecheckthelevelprobesof
the liquid suction tubes (detergent, neutraliser and/or oth-
er): pull one tube at a time out of the container and check if
thecorrespondingalarmashesonthedisplay,indicating
theneedforliquidrelling.
7) After running 3-4 washing tests, clean the water lters
placed in the lower compartment of the machine (Fig. 9).
8)
Check that the drain trap does not show water leaks and is
rmlysecuredtoboththemachineandthedrain.
9) Check if the pumps correctly suck washing liquids. To do
this, check that the liquid rises along the tube connected to
them.
Fig. 13 - 1) ON-OFF switch. - 2) USB pendrive socket for PC
connection and to use the «Win Tuttnauer Service tool wash»
program. - 3) USB pendrive socket for inserting or extracting
the pre-lled programs. - 4) Rubber cap to close access to the
emergency release.