44
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
5) Start the machine using the ON-OFF switch (1 Fig. 13).
6) Atthebeginningoftherstcyclecheckthelevelprobesof
the liquid suction tubes (detergent, neutraliser and/or oth-
er): pull one tube at a time out of the container and check if
thecorrespondingalarmashesonthedisplay,indicating
theneedforliquidrelling.
7) After running 3-4 washing tests, clean the water lters
placed in the lower compartment of the machine (Fig. 9).
8)
Check that the drain trap does not show water leaks and is
rmlysecuredtoboththemachineandthedrain.
9) Check if the pumps correctly suck washing liquids. To do
this, check that the liquid rises along the tube connected to
them.
Fig. 13 - 1) ON-OFF switch. - 2) USB pendrive socket for PC
connection and to use the «Win Tuttnauer Service tool wash»
program. - 3) USB pendrive socket for inserting or extracting
the pre-lled programs. - 4) Rubber cap to close access to the
emergency release.
1
2
3
4