EasyManua.ls Logo

Tuttnauer TIVA2 - 6 Using the Machine; Commissioning Instructions

Tuttnauer TIVA2
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42
Instructions for use - Tiva2 - Tiva2H Thermodisinfector
6. Using the machine
Before starting the machine, the operator in charge must have
read and understood this whole manual, in particular the infor-
mation given in section «3 Safety and prevention».
Furthermore, before starting work, check that the machine is
in order and that all parts subject to wear and deterioration are
fullyefcient.
6.1 Commissioning instructions
These control operations (reported below) are per-
formed to check if the machine works properly and
should be performed when the machine installation is
completed.
1) Open the tap that supplies water to the machine.
WARNING. The water should NOT ow into the tank;
otherwise the water loading solenoid valves (in the low-
er compartment of the machine) are dirty or blocked
due to long storage in the warehouse and therefore they
must be cleaned.
Check that there are no water leaks in the pipe ttings.
2) Check that the suction tubes (located in the lower part of
the machine) within the corresponding containers/tanks
containing the liquids provided (detergent, neutralising,
and/or others), are correctly inserted.
3) Using the main circuit breaker, supply power to the ma-
chine.
4) Check that the water supply exible hoses are properly
connected.

Related product manuals