Instructions for use - Tiva2 - Tiva2H Thermodisinfector
- Check the electrical safety according to VDE 0701/0702.
- Filterinthewaterowvalves.
- Replacetheprelter.
- ReplacetheHEPAlter.
- Check the dosing systems.
- Thermostat sensor, thermoelectric measurements.
- Conditions and functionality of the spray arms.
- Safety thermostat.
- Door gasket.
- Steam condenser.
- The functionality of the drain pump.
- Pressure switch.
- Drain pipes.
- Waterowpipes.
7.1.3 Air suction lter
Thethermodisinfectorisstandardequippedwithasuctionlter
for drying air (Fig. 19), to be replaced after about 100 hours of
work (equal to about 200 cycles).
The machine can also be equipped with an «absolute» «HEPA
H14»classsupplementaryairlter(onrequest),incompliance
with EN 1822.
We recommend replacing the HEP lter every 500 working
hours, which is equivalent to 1000 operating cycles circa.
Fig. 19 - Air suction lter